Thyrotoxicosis antidote assay along with concurrent medication; chromatographic and environmental issues
BMC Chemistry volume 19, Article number: 124 (2025)
Abstract
A thionamide medication, Methimazole (MTZ), is a crucial antidote of thyroid hormone in cases of toxic nodular goiter or thyrotoxicosis by decreasing thyroid hormone synthesis. Propranolol (PRP), a beta blocker, is commonly co-administered with MTZ for controlling tachycardia associated with hyperthyroidism. Quantitative determination and tracing of MTZ in plasma in the presence of its co-administered medication, like PRP, is of paramount importance due to the reported toxic manifestations related to MTZ long-term administration that include agranulocytosis, hepatitis, arthritis, and skin rashes. An environmentally benign FDA-validated TLC densitometric approach was established for the first time for simultaneous and quantitative analysis of MTZ and PRP in pure form and spiked human plasma. The work is considered a mimetic study for their co-administration. Successful resolution between them was achieved on Merck® silica gel plates using a mixture of ethyl acetate: acetone: 33% NH3 solution in a ratio of (9: 1: 0.05, by volume) as a developing phase and UV scanning at 254 nm. Adding hydrocortisone acetate (HCA) as an internal standard eliminated the matrix effect variation. Reasonable resolution of the developed peaks was attained, with Rf values of 0.07, 0.19, 0.67, and 0.81 for plasma, PRP, MTZ, and HCA, respectively. Four greenness and viability rating approaches were applied to check and measure the greenness aspects of the suggested method toward the eco-system, and the outcomes were convenient and satisfactory. Also, the verification domains were tested using US-FDA bio-analytical specifications where reliable and acceptable outcomes were obtained with satisfied % recoveries for the quality control samples ranging from (100.39–104.44%) and (96.01–100.72%) for MTZ and PRP, respectively, with low RSD values indicating good accuracy and precision of the proposed method.
Introduction
Methimazole (MTZ), Fig. 1, is a thionamide medication used for managing hyperthyroidism, which is a medical condition characterized by over-production of thyroid hormone (thyrotoxicosis), leading to health complications [1]. These complexities include atrial fibrillation (with its consequent complications such as stroke and heart failure), in addition to osteoporosis, hard conceiving, miscarriage, menstrual disturbances, muscle weakness, and Graves'disease, which is associated with light sensitivity and double blurred vision [2]. MTZ works by suppressing the thyro-peroxidase enzyme, which is responsible for thyroid hormones formation, leading to the inhibition of thyroid hormones synthesis, including thyroxine (T4) and triiodothyronine (T3), which results in restoring normal thyroid function [1].
Propranolol (PRP), Fig. 1, is a β-adrenoceptor antagonist [3]. It is used as a treatment for controlling hypertension, myocardial infarction, arrhythmia, and angina. It is also used to manage hyperthyroidism, anxiety disorders, and hyperthyroid tremor in case of sympathetic overactivity cases [4].
PRP and MTZ are co-administered for hyperthyroidism treatment, as MTZ can optimize thyroid hormone production and enhance hepatic function. At the same time, PRP can adjust the heart rate and improve the health quality of the patient, with the lowest side effects [5]. Unfortunately, long-term administration of MTZ was found to be related to agranulocytosis, toxic hepatitis, arthritis, and skin rashes symptoms [1].
In the literature, MTZ was analyzed and determined using spectroscopic [6, 7], TLC [8], HPLC [9,10,11,12], GC [13], and voltammetric [14, 15] methods in different samples and conditions. In addition, PRP was analyzed and determined using spectroscopic [16,17,18,19], TLC [20, 21], HPLC [22,23,24,25,26], and voltammetric [27, 28] methods in different samples and conditions.
MTZ and PRP had not been determined simultaneously in any published work. Such toxic manifestations created an increasing demand for developing an accurate and sensitive assay method for MTZ quantification along with PRP, which improves its performance, especially in case of a lack of any published analytical approach for their simultaneous determination upon review of the literature. Therefore, the purposes of this article are to develop an environmentally benign TLC assay method for quantifying and separating PRP and MTZ in pure form and spiked human plasma, to verify the suggested method in compliance with US-FDA specifications, and to assess the proposed method's environmental impact and viability by applying various evaluating tools.
In addition, four greenness and viability rating approaches were applied to check and assess the greenness aspects of the suggested method toward the eco-system, namely: eco-scale assessment (ESA) [29], green analytical procedure index (GAPI) [30], analytical greenness metric approach (AGREE) [31], and practicability metric approach (BAGI) [32]. Also, the whiteness of the method was appraised using the red–green–blue (RGB) Excel sheet [33]. The outcomes were convenient and satisfactory, showing that the proposed TLC method is acceptably green and eco-friendly to the environment and human lives.
Experimental
Instruments
For TLC—densitometric approach
-
TLC (Linomat V) applicator was used to apply samples with a 100 µL Camag (Muttenz, Switzerland) syringe.
-
The detection of the resulting peaks was achieved with the assessment of a TLC densitometer scanner (Camag, Switzerland) by adjusting the following parameters (the mode of scanning: absorbance, the slit measures: 6 × 0.45 mm, the scanning speed: 20 mm/s, and the radiation source: a deuterium lamp.
-
The software used was WinCATS (Camag, version 3.15), which was connected to a TLC scanner to record and manipulate the resulting data.
-
Merck® Silica gel-coated aluminium plates 60 F254 (25 × 10–2 mm thickness) served as stationary phase, and the plate dimensions were (20 × 20 cm).
For plasma sample preparation
-
Digital balance (Sartorius, German) was used to withdraw the accurate weight of the sample powders.
-
Different sample volumes were withdrawn using Rongtai micropipette devices with various volumes of 0.1–100 µL (Shanghai, China).
-
An ultra-sonicator (Sonix TV SS-serious) was used to sonicate the samples (New York, USA).
-
A Low-speed 80–2 C electric centrifuge (4000 rpm) with 12 × 20 mL capacity and 110 v/220v power supply (Zjmzym, China) was used to centrifuge the prepared samples.
-
Syringe filters with 45 × 10–2 µm pore size (Npore Ghaziabad, India) were utilized to filter the samples.
-
A vortex mixer (Hwashin, Seoul, Korea) was employed to mix the samples.
Materials
Pure standards
-
PRP and hydrocortisone acetate (HCA) with certified purities of (99.21 and 99.63%, respectively) were provided by Al-Kahira Company for Pharmaceuticals (Giza, Egypt).
-
MTZ, with a certified purity of 99.53, was bought from Sigma Aldrich (Cairo, Egypt)
Chemicals and reagents
-
Methanol (Fischer, UK) of HPLC grade was utilized.
-
Ethyl acetate, acetone, and 33% NH3 solution of analytical grade were bought from AL-Nasr Pharmaceutical and Chemical Company (Cairo, Egypt).
-
No prior purification was employed for the chemicals and solvents utilized.
Standard solutions and blank plasma sample
-
Stock solutions of 1000 µg/mL of PRP, MTZ, and HCA were made separately in three 10-mL volumetric flasks, using methanol as solvent.
-
Further working solutions of 100 µg/mL and 10 µg/mL of PRP and MTZ were made separately from the previously made stock solutions into two different 10-mL volumetric flasks using the same solvent.
-
For the blank plasma samples, after transferring 1 mL of drug-free plasma into a 10-mL measuring flask, methanol was added to adjust the volume to 10 mL. Also, plasma proteins were precipitated before collecting the clear supernatant, and the sample was centrifuged.
Procedures
Chromatographic parameters
Samples were applied to the TLC plates using a Camag Linomat IV applicator in bands (three mm wide) spaced six mm from each other and ten mm apart from the bottom border. The mobile system was made up of ethyl acetate: acetone: 33% NH3 solution (9: 1: 0.05, by volume), which was put into a jar made of glass and left for saturation for 10 min. After that, the plates were put into the jar, tightly sealed, and left until the mobile system reached the plate's front line, one centimeter in front of the plate's end. In the end, UV scanning was employed on the dry plates to detect the resulting peaks spectrophotometrically at 254 nm. The samples were applied with a volume of 10 µL.
Calibration graphs construction and preparation of quality control samples
For pure samples
Using their respective stock solutions, two groups of 10-mL volumetric flasks were filled with varying concentrations between (0.04–1.0) µg/band for PRP and (0.1–1.0) µg/band for MTZ. Each flask received 0.1 mL of the internal standard (HCA) solution (1 mg/mL) before the volume was marked with methanol. Subsequently, 10 µL injections of each sample were applied in triplicate, and the procedures listed under"chromatographic conditions"were employed. Ultimately, calibration curves were created by plotting the integrated peak area ratios (peak area of the analyte/peak area of internal standard) against the corresponding concentration in µg/band.
For spiked human plasma samples
In the limits of (0.06–1.0) and (0.1–1.2) µg/band of PRP and MTZ, respectively, various concentrations were created utilizing their corresponding stock and working solutions into two sets of 10-mL measuring flasks. Each flask received 0.5 mL of internal standard (HCA) stock solution and 1 mL of drug-free plasma, and then the volume was completed with methanol. After 5 min sonication of the samples, 5 min centrifugation was performed to exclude the precipitated plasma proteins, and the clear supernatant was applied in 10 µL volume on the TLC plates as triplicates following the previously outlined"chromatographic conditions."At the end, the calibration curves were computed by graphing the integrated peak area ratios (peak area of the analyte/peak area of internal standard) of the spiked human plasma samples and the corresponding concentration in µg/band. In addition, as FDA specifications stated, four quality control samples should be prepared and analyzed employing a similar procedure for both drugs [34].
Results and discussion
Thin-layer Chromatography is a reliable lab method for separating and examining different components in a mixture. Its various benefits include taking little in the way of tools and reagents, making it reasonably simple and inexpensive. TLC provides rapid results, allowing the speedy identification and separation of mixture components. It is also suitable for rare or precious samples because it only requires small sample volumes. Additionally, TLC can easily be applied to both qualitative and quantitative evaluations. Its adaptability in disentangling various medications and constituents from intricate matrices amplifies its usefulness in experimental settings [35,36,37,38]. Therefore, this study offers a newly verified TLC method for simultaneous analysis and quantification of MTZ, which is reported to be toxic upon long-term administration [1], in addition to PRP, which synergizes its action in their binary mixture and spiked human plasma. It has the benefits of being environmentally green, quick, easy, inexpensive, and requiring little amounts of samples and reagents.
Method optimization
The selection of the developing phase composition and ratio
Different developing solvents were tested to separate MTZ and PRP from plasma for optimum selectivity and resolution. Different developing solvents with different ratios were tested: methanol (MTH): acetone (AC), ethyl acetate (EA): AC and EA: MTH in the ratios of 6: 4, 4: 6, 7: 3, 3: 7, 9: 1, 1: 9. A developing system of MTH: AC (6: 4, by volume), resulted in separated PRP and MTZ peaks with tailed PRP one. Another trial was carried out by replacing methanol with EA; as greener than it [39], a developing system of EA: AC (6:4, by volume) was used that improved PRP peak symmetry, but the MTZ band appeared near the solvent front. Increasing EA proportion to be EA: AC (9: 1, by volume) attained the best peaks'symmetry, with MTZ peak away from the solvent front, but PRP'one appeared near the baseline adjacent to the plasma band. At this step, different drugs (hydrocortisone, HCA, glibenclamide, and etoricoxib) were tested to be used as internal standards, applying the last developing system. Etoricoxib and hydrocortisone bands appeared at Rf values close to that of MTZ. At the same time, GLB one was near baseline, and HCA appeared near the front line but separated from the MTZ peak, so it was chosen as an internal standard, as shown in Fig. 2. Finally, ammonia solution (33%) was added to the system (EA: AC, NH3, 9: 1: 0.05, by volume), led to good separation of PRP peak away from that of plasma and the baseline and achieved reasonable separation between MTZ and HCA away from front line while attaining optimum peaks symmetry. Accordingly, the final developing system was EA: AC: 33% NH3 solution in a ratio of (9: 1: 0.05, by volume).
The selection of the detection wavelength
Different scanning trials used other wavelengths, including 230, 254, and 260 nm. The optimum detection wavelength was 254 nm, with the best sensitivity and lowest noise.
The selection of the internal standard
The chromatographic separation and performance were optimum after trying different drugs as internal standards, such as hydrocortisone, HCA, glibenclamide, and etoricoxib. They attained the best results when using HCA as an internal standard. It was the best one for compensating for changes in extraction efficiency, and its peak shape was symmetric and separated with good resolution from the peaks of other components under the same conditions.
Slit dimensions of scanning light beam
The band dimensions were tested and enhanced by optimizing the slit dimensions of the scanning light beam. The interference of the adjacent band should be eliminated. Different trials of slit dimensions were carried out. The slit dimension of (6 × 0.45) mm was found to be the optimum one with the best sensitivity. Finally, the studied drugs were well separated at RF values of 0.07, 0.19, 0.62, and 0.81 for plasma, PRP, MTZ and HCA, respectively using ethyl acetate: acetone: 33% NH3 solution in a ratio of (9: 1: 0.05, by volume) as a developing system, HCA as internal standard, 254 nm as UV-scanning wavelength and slit dimension of (6 × 0.45) mm, as shown in Fig. 3.
2D chromatogram of a blank human plasma, b plasma and 0.5 µg/band of the internal standard hydrocortisone acetate, c spiked human plasma with 0.8 µg/band of propranolol, 1.0 µg/band of methimazole and 0.5 µg/band of the internal standard hydrocortisone acetate, using ethyl acetate: acetone: 33% NH3 solution (9: 1: 0.05, by volume) as a developing system at 254 nm
Method development
Plotting the peak area ratio (peak area of the drug/peak area of IS) against the corresponding concentration in µg/band allowed for constructing the calibration curves, which were then created following the procedures outlined in the"chromatographic conditions"section. As illustrated in Fig. 4, linear regressions were seen in the concentration ranges of 0.04–1.0 µg/band for PRP and 0.1–1.0 µg/band for MTZ. Table 1 provides the criteria for the regression equations. As seen in Fig. 3c, the obtained Rf values for plasma, PRP, MTZ, and HCA were 0.07, 0.19, 0.62, and 0.81, respectively.
Calibration curves relating the peak area ratios of a propranolol in pure form to its concentrations in the range 0.04–1.0 μg/band, b methimazole in pure form to its concentrations in the range 0.1–1.0 μg/band, c propranolol in spiked human plasma to its concentrations in the range 0.06–1.2 μg/band, d methimazole in spiked human plasma to its concentrations in the range 0.1–1.2 μg/band using the proposed TLC method
Method validation
The developed bio-analytical TLC method was validated by adhering to FDA [34] as follows:
Calibration curves, lower and upper limits of quantification (LLOQ & ULOQ)
For PRP and MTZ, linear calibration curves were produced with concentration ranges of 0.06–1.2 µg/band and 0.1–1.2 µg/band, respectively, after plasma samples spiked with the medications under study. As indicated in Table 1, the LLOQ and ULOQ were detected to be 0.06 and 1.2 µg/band for MTZ and 0.1 and 1.2 µg/band for PRP.
Accuracy, precision, and quality control samples (QCs)
LLOQ, LQC, MQC, and HQC were the four QC samples validated to assess the established procedure's precision and accuracy. For PRP, (0.06, 0.2, 0.4, and 1.0) µg/band and (0.1, 0.3, 0.6, and 1.0) µg/band for MTZ were the chosen QC concentrations. The concentrations of the medications under study were calculated using the matching regression equations provided in Table 1, and the results were deemed acceptable, as indicated in Table 2.
Specificity and selectivity
By comparing the TLC chromatograms produced from the application of drug-free plasma samples (from six different plasma batches) and spiked plasma samples containing the two investigated medications at their LLOQ and IS, the specificity of the sample was verified. The TLC chromatograms in Fig. 3 demonstrated the new method's selectivity, displaying a good separation between plasma, PRP, MTZ, and the internal standard HCA without interference from the plasma matrix and plasma proteins.
The freeze–thaw and benchtop stability
The stability of the drugs under study was assessed using three freeze–thaw cycles and benchtop stabilities, and the results were acceptable and satisfactory. They demonstrated the stability of the drugs under study, as indicated in Table 3.
The extraction recovery
The extraction recovery for the proposed drugs was calculated and analyzed by matching the areas beneath the peaks of the extracted samples at LQC, MQC, and HQC concentrations versus extracts of blanks spiked with the analyte post-extraction at the same concentration levels [34]. The results in Table 4 confirmed that the current work's extraction method was accurate.
System suitability parameters
The resolution, tailing, and selectivity factors were among the many metrics calculated to assess the system's suitability. The results were validated per the permitted ranges, as shown in Table 5 [40].
Robustness
By making minor, carefully considered adjustments to several parameters, including the developing system proportions, the scanning wavelength, and the saturation period, the robustness of the devised TLC method was examined, and the findings were recorded in Table 1.
Greenness assessment of the developed TLC method
Eco-scale assessment (ESA)
The analytical eco-scale is one of the authorized metrics for assessing the process's environmental friendliness [29]. Table 6 displays the eco-scale ratings determined for the established TLC. The findings demonstrate that, per the Globally Harmonized System of Classification and Labeling of Chemicals (GHS), the established approach got an eco-scale grade of 80, explaining that the suggested TLC method is acceptably green [29] with minimum impact on the environmental life and human health.
Green analytical procedure index (GAPI)
A reliable resource that offers an extensive environmental assessment of the entire analytical methodology, from the collection of the sample to the final results, is the GAPI tool [30]. Table 6 presents an evaluation and illustration of the five primary GAPI pictogram sections of the proposed TLC technique, composed of four green, nine yellow, and two red segments. The results proved that the proposed method is acceptably green with minimum impact on environmental life and human health.
Analytical greenness metric approach (AGREE)
The AGREE tool, an efficient online method for evaluating green characteristics [31], can be downloaded from a defined online link listed in the AGREE research [31]. Table 6 illustrates and displays the colored pictogram and total AGREE score for the developed TLC method, which is equal to 0.73 with a light green shade, which indicates the greenness of the developed TLC method and that it is an eco-friendly analytical method to the environment and human health.
Practicability metric approach (BAGI)
The analytical method's performance is assessed by the BAGI tool [32]. This tool considers ten attributes related to the technique, the instrument, and the samples. A score and a pictogram are produced. In our method, several samples can be simultaneously analyzed within an hour; a simple extraction procedure, a semi-automated instrument, and commercially available solvents were used. The size of each sample was 5 mL, containing 1 mL of plasma. As displayed in Table 6, a pictogram and a score of 75 were produced with a light blue shade, which indicates the greenness of the developed TLC method and that it is an eco-friendly analytical method for the environment and human health.
Whiteness assessment
The Whiteness assessment of a method was done by assessing three aspects [41,42,43,44,45], including execution, greenness, and workability. In the red–green–blue (RGB) Excel sheet downloaded from the original article describing the assessment method [33], several parameters are given scores according to the degree of compliance with the required values. The method achieved the scope of the application, had low LOD and LOQ values, and had acceptable accuracy and precision, so it got a score of 100 for the execution aspect. The total number of pictograms for the used solvents was 9, which wasn't too high; the waste amount was over 100 mL that wasn't treated; the energy consumption was moderate, and the occupational hazards were low, so the greenness score was 90. Finally, the method was cost-, time-, and energy-efficient; the procedure was simple, although it could not be miniaturized. A score of 95 was given for workability. The overall score of whiteness was 95, as found in Table 6.
A brief comparison between the developed and the reported TLC methods
The developed TLC method was compared briefly with the previously published chromatographic methods for determining PRP and MTZ in plasma matrices. As illustrated in Tables 7, 8 the proposed method has advantages over the previously published methods of determining PRP and MTZ in spiked human plasma. It is more accurate than the reported methods [11,12,13, 21, 25, 26] for estimating PRP and MTZ. In addition, the proposed method is more precise than the reported ones [11, 12, 26] for determining the two drugs. Moreover, the proposed method used greener mobile phase solvents than the reported methods [11, 13, 21, 25, 26], which used chloroform and acetonitrile (toxic and non-green), proving that the developed method is greener and safer for the environment and human health. On the other hand, the proposed method has the drawback of being less sensitive than the reported methods [11,12,13, 21, 25, 26] for the determination of PRP and MTZ.
Conclusion
Using HCA as an internal standard, a sensitive green TLC technique was created to simultaneously measure PRP and MTZ in their combination and spiked human plasma. For MTZ, being toxic upon long-term usage and being synergized with PRP created an urgent demand for developing such an assay. The process has the benefits of being quick, easy, affordable, and environmentally friendly, and it requires a small number of samples and reagents. All validation parameters met the US-FDA acceptance criteria for verifying bio-analytical methods. The newly suggested TLC approach is environmentally friendly. It has little to no negative impacts on human or natural life, according to a greenness evaluation conducted using three different tools.
Data availability
The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.
References
Awosika AO, Singh G, Correa R. Methimazole. National library of medicine. 2023. https://www.ncbi.nlm.nih.gov/books/NBK545223/. Accessed 13 Sept 2023.
Ross DS, Burch HB, Cooper DS, Greenlee MC, Laurberg P, Maia AL, et al. American thyroid association guidelines for diagnosis and management of hyperthyroidism and other causes of thyrotoxicosis. Thyroid. 2016;26:1343–421.
The British Pharmacopoeia. Medicines and healthcare products regulatory agency. London: UK; 2013.
Al-Majed AA, Bakheit AH, Aziz HAA, Alajmi FM, AlRabiah H. Propranolol. In Profiles of drug substances, excipients and related methodology. Academic Press. 2017; 42: 287–338.
Lu S, Shi D. Influence of propranolol plus methimazole on curative efficacy and thyroid function of patients with hyperthyroidism. Am J Transl Res. 2024;16:1375–82.
Dong C, Zhang Y, Guo L, Li QM. Spectrophotometric determination of methimazole in pharmaceutical, serum and urine samples by reaction with potassium ferricyanide-Fe (III). J Anal Chem. 2010;2023(65):707–12.
Skowron M, Ciesielski W. Spectrophotometric determination of methimazole, D-penicillamine, captopril, and disulfiram in pure form and drug formulations. J Anal Chem. 2011;66:714–9.
Mc D, Pardeshi KK. Stability indicating chromatographic method for estimation of methimazole. Int J Pharm Biol Sci. 2019;9:843–8.
Zakrzewski R. Determination of methimazole in pharmaceutical preparations using an HPLC method coupled with an iodine-azide post-column reaction. J Liq Chromatogr Relat Technol. 2008;32:383–98.
Kuśmierek K, Bald E. Determination of methimazole in urine by liquid chromatography. Talanta. 2007;71:2121–5.
Meulemans A, Manuel C, Ferriere C, Valpillat M. Determination of methimazole in plasma by high performance liquid chromatography. J Liq Chromatogr. 1980;3:287–98.
Ito Y, Sakai H, Nagao H, Suzuki R, Odawara M, Minato K. Development of a high-performance liquid chromatography–tandem mass spectrometric method for the determination of methimazole in human blood matrices. J Chromatogr B. 2020;1144: 122083.
Floberg S, Lanbeck K, Lindström B. Determination of methimazole in plasma using gas chromatography—mass spectrometry after extractive alkylation. J Chromatogr B: Biomed Sci App. 1980;182:63–70.
Aslanoglu M, Peker N. Potentiometric and voltammetric determination of methimazole. J Pharm Biomed Anal. 2003;33:1143–7.
Shahrokhian S, Ghalkhani M. Voltammetric determination of methimazole using a carbon paste electrode modified with a schiff base complex of cobalt. Electroanalysis. 2008;20:1061–6.
Doshi AK, Patel BN, Patel CN. Development and validation of spectrophotometric method for simultaneous determination of propranolol hydrochloride and flunarizine dihydrochloride in their combined dosage formulation. Int J Pharm Sci Res. 2012;3:1741.
El-Emam AA, Belal FF, Moustafa MA, El-Ashry SM, El-Sherbiny DT, Hansen SH. Spectrophotometric determination of propranolol in formulations via oxidative coupling with 3-methylbenzothiazoline-2-one hydrazone. Il Farmaco. 2003;58:1179–86.
El-Didamony AM. A sensitive spectrophotometric method for the determination of propranolol HCl based on oxidation bromination reactions. Drug Test Anal. 2010;2:122–9.
Sharma DK, Singh J, Raj P. Spectrophotometric determination of propranolol hydrochloride and metoprolol tartrate in pharmaceutical dosage forms, spiked water and biological fluids. Int J Pharm Pharm Sci. 2018;10:107–15.
Bhavar G, Chatpalliwar VA. Quantitative analysis of propranolol hydrochloride by high performance thin layer chromatography. Indian J Pharm Sci. 2008;70:395.
Mennickent S, Vega M, Vega H, Diego M, Fierro R. Quantitative determination of propranolol in human serum by high-performance thin-layer chromatography. J Planar Chromat. 2012;25:54–9.
Torul H, Tamer U. Determination of enantiomers of atenolol and propranolol in pharmaceutical formulations by HPLC. J AOAC Int. 2011;94:833–8.
Morante-Zarcero S, Sierra I. Simultaneous enantiomeric determination of propranolol, metoprolol, pindolol, and atenolol in natural waters by HPLC on new polysaccharide-based stationary phase using a highly selective molecularly imprinted polymer extraction. Chirality. 2012;24:860–6.
Shabir GA. Development and validation of RP-HPLC method for the determination of methamphetamine and propranolol in tablet dosage form. Indian J Pharm Sci. 2011;73:430.
Srikanth MV, Ram BJ, Sunil SA, Rao NS, Murthy KV. Development and validation of HPLC method for estimation of propranolol HCl in human plasma. J Sci Ind Res. 2012;71:120–3.
Salman SAB, Sulaiman SA, Ismail Z, Gan SH. Quantitative determination of propranolol by ultraviolet HPLC in human plasma. Toxicol Mech Methods. 2010;20:137–42.
Sartori ER, Medeiros RA, Rocha-Filho RC, Fatibello-Filho O. Square-wave voltammetric determination of propranolol and atenolol in pharmaceuticals using a boron-doped diamond electrode. Talanta. 2010;81:1418–24.
Oliveira GG, Azzi DC, Vicentini FC, Sartori ER, Fatibello-Filho O. Voltammetric determination of verapamil and propranolol using a glassy carbon electrode modified with functionalized multiwalled carbon nanotubes within a poly (allylamine hydrochloride) film. J Electroanal Chem. 2013;708:73–9.
Gałuszka A, Migaszewski ZM, Konieczka P, Namieśnik J. Analytical eco-scale for assessing the greenness of analytical procedures. TrAC Trends Anal Chem. 2012;37:61–72.
Płotka-Wasylka J. A new tool for the evaluation of the analytical procedure: green analytical procedure index. Talanta. 2018;181:204–9.
Pena-Pereira F, Wojnowski W, Tobiszewski M. AGREE—analytical GREEnness metric approach and software. Anal Chem. 2020;92:10076–82.
Manousi N, Wojnowski W, Płotka-Wasylka J, Samanidou V. Blue applicability grade index (BAGI) and software: a new tool for the evaluation of method practicality. Green Chem. 2023;25:7598–604.
Mateusz Nowak P, Wietecha-Posłuszny R, Pawliszyn J. White analytical chemistry: an approach to reconcile the principles of green analytical chemistry and functionality. TrAC, Trends Anal Chem. 2021;138: 116223.
US Department of Health and Human Services, Food and Drug Administration Center for Drug Evaluation and Research (CDER), Center for Veterinary Medicine (CVM). Bioanalytical method validation guidance for industry; 2018.
Abdelhamid NS, Magdy MA, Anwar BH, Farid NF. US FDA-validated TLC method with four greenness assessment evaluations for simultaneous determination of prednisolone and esomeprazole in spiked human plasma. Biomed Chromatogr. 2022;36: e5343.
Farid NF, Abdelhamid NS, Naguib IA, Anwar BH, Magdy MA. Quantitative determination of anti-migraine quaternary mixture in presence of p-aminophenol and 4-chloroacetanilide. J Chromatogr Sci. 2022;60:538–44.
Farid NF, Naguib IA, Abdelhamid NS, Anwar BH, Magdy MA. Validated ecofriendly chromatographic method for quantitative determination of anti-migraine quaternary mixture. J Sep Sci. 2020;43:2330–7.
Abdelhamid NS, Naguib IA, Anwar BH, Magdy MA. A validated HPTLC method for the quantitative determination of duloxetine hydrochloride and 1-naphthol in bulk and pharmaceutical formulation. J Planar Chromatogr. 2020;33:391–6.
Elzanfaly ES, Hegazy MA, Saad SS, Salem MY, Abd El Fattah LE. Validated green high‐performance liquid chromatographic methods for the determination of coformulated pharmaceuticals: a comparison with reported conventional methods. J Sep Sci. 2015; 38: 757–763.
Srivastava M. High-performance thin-layer chromatography (HPTLC). Springer Science & Business Media. London: UK; 2010.
Prajapati P, Patel K, Shakar Pulusu V, Haque A, Abul Kalam M, Shah S. Microwave-aided sensitive and eco-friendly HPTLC method for estimation of mirabegron using NBD-Cl as bio-sensing fluorescent probe by Integration of principal component analysis, response surface modelling and white analytical chemistry approach. Microchem J. 2024;207: 111862.
Prajapati P, Shahi A, Acharya A, Pulusu VS, Shah S. Implementation of white analytical chemistry–assisted analytical quality by design approach to green liquid chromatographic method for concomitant analysis of anti-hypertensive drugs in human plasma. J Chromatogr Sci. 2023. https://doiorg.publicaciones.saludcastillayleon.es/10.1093/chromsci/bmad054.
Prajapati P, Pulusu VS, Shah S. Red, green, and blue model-based assessment and principles of white analytical chemistry to robust stability-indicating chromatographic estimation of thiocolchicoside and diclofenac sodium. J AOAC Int. 2023;106:1374–88.
Prajapati P, Pulusu VS, Shah S. White analytical chemistry-driven stability-indicating concomitant chromatographic estimation of thiocolchicoside and aceclofenac using response surface analysis and red, green, and blue model. J Sep Sci. 2023;46:2300139.
Prajapati P, Pulusu VS, Shah S. Principles of white analytical chemistry and design of experiments to development of stability-indicating chromatographic method for the simultaneous estimation of thiocolchicoside and lornoxicam. J AOAC Int. 2023;106:1654–65.
Acknowledgements
The authors thank the Faculty of Pharmacy, Beni-Suef University, for supporting us during this work.
Funding
Open access funding provided by The Science, Technology & Innovation Funding Authority (STDF) in cooperation with The Egyptian Knowledge Bank (EKB).
Author information
Authors and Affiliations
Contributions
Raghda A. Emam: Formal analysis, Writing—original draft, Aml. A. Emam: Writing—review and editing, Methodology, Rehab M. Abdelfatah: Writing—review and editing, Supervision, Bassma H. Hossam: Writing—original draft, Supervision.
Corresponding author
Ethics declarations
Ethics approval and consent to participate
This work has been approved by the research ethics committee of the Beni-Suef University, Faculty of Pharmacy, under serial number REC-H-PhBSU-24019. Informed consent was acquired from every participant. This study complies with the Helsinki Declaration.
Competing interests
The authors declare no competing interests.
Additional information
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.
About this article
Cite this article
Emam, R.A., Emam, A.A., Abdelfatah, R.M. et al. Thyrotoxicosis antidote assay along with concurrent medication; chromatographic and environmental issues. BMC Chemistry 19, 124 (2025). https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s13065-025-01487-1
Received:
Accepted:
Published:
DOI: https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s13065-025-01487-1