Skip to main content

Table 2 Intra and inter-assay precision and accuracy of LLOQ, LQC, MQC, and HQC of propranolol and methimazole in spiked human plasma samples

From: Thyrotoxicosis antidote assay along with concurrent medication; chromatographic and environmental issues

Component

Concentration (µg/band)a

Intra-day

Inter-day

Recovery %

RSD %

Bias %b

Recovery %

RSD %

Bias %b

PRP

LLOQ

0.06

103.61

2.090

3.61

102.83

2.971

2.83

LQC

0.2

104.44

3.990

4.44

98.72

4.018

− 1.28

MQC

0.4

100.39

1.890

0.39

102.54

3.111

2.54

HQC

1.0

103.30

1.040

3.30

99.28

2.882

− 0.82

MTZ

LLOQ

0.1

96.01

1.798

− 3.99

102.24

3.178

2.24

LQC

0.3

96.65

2.820

− 3.35

104.85

4.281

4.85

MQC

0.6

98.35

1.987

− 1.65

102.07

2.753

2.07

HQC

1.0

100.72

3.059

0.72

97.21

4.071

− 2.79

  1. a Average of 3 experiments
  2. b Bias = [(measured concentration—nominal concentration)/nominal concentration] × 100