A stability-illustrating HPLC-DAD method for assessment of two veterinary anti-parasitic drugs: appraisal of the method’s greenness and blueness
BMC Chemistry volume 19, Article number: 35 (2025)
Abstract
This paper represents an effective and reliable high-performance liquid chromatography-diode array detector (HPLC-DAD) method for the regular assay of Clorsulon (CLR) and Moxidectin (MOX) anti-parasitic drugs in injection solution and pure powder without derivatization processes. The mobile phase was composed of acetonitrile: methanol: water: acetic acid (56.0: 36.0: 7.5: 0.5 by volume). Besides, a Supelcosil C18 (4.60 mm ×15.0 cm, 5.0 μm) column was selected for completing the separation and quantitation of the two aforementioned veterinary drugs at a wavelength of 254 nm. The flow rate was set at 2.0 mL min−1 at the isocratic approach. We have conducted the degradation experiments using the HPLC-DAD instrument, adhering to the guidelines of the International Conference of Harmonization (ICH), subjecting CLR and MOX to light, heat, basic, acidic, and oxidative stressful conditions to figure out the ideal storage conditions and the possible medications that can be co-administered with them. CLR and MOX were quantified linearly from 400.0 to 1200.0 and from 40.0 to 120.0 µg mL−1, respectively. The maximum recorded degradation results were in acidic, basic, and oxidative conditions. Therefore, strong basic or acidic medications and oxidants shouldn’t be combined with CLR and MOX in a co-formulated medication. Greenness, carbon footprint, and blueness assessments for the novel method were conducted to verify the sustainability and functionality. The thirteen subdivisions in the GAPI pictogram, which were categorized as either green or yellow, refer to moderate green aspects. The final AGREE score of 0.56 and the majority of its subdivisions, ranging from dark green to yellow, indicated a relatively moderate level of greenness. This was primarily due to the significant acetonitrile content (56%) in the mobile phase. Using the HEXAGON method, the ultimate score is 0 out of 5 since the total calculated carbon footprint is less than 0.10. An eco-friendly method is one with a reduced carbon footprint score. The innovative HPLC method’s functioning and utility are indicated via its overall BAGI score of 80.0. Generally, the outcomes of the AGREE and GAPI pictograms indicate that the HPLC-DAD has a greenness feature, despite its moderate sensitivity.
Graphical Abstract

Introduction
For precise drug quantification free from interference from any contaminants and/or degradations, an effective stability-indicating method must be developed [1,2,3]. Implementing an environmentally friendly and sustainable analytical technique is also important for maintaining a clean environment for our world and the health of scientists [4,5,6,7]. These days, there is a lot of interest in sustainability and green analytical chemistry. Techniques based on sustainable chromatography are frequently employed in the management of contaminants found in water, air, and soil [8, 9]. Applications for sustainability are also expanded to include the potential of extracting solvents being reused before chromatographic analysis and the decrease of extracting solvents [10,11,12,13]. It is greatly welcomed when sustainability principles are applied to chromatography-based procedures to guarantee both operator safety and environmental preservation. These environmentally friendly guidelines must be followed in their entirety [6]. Thus, the analyst should bear in mind that the quality and applicability of the developed method will be enhanced by the use of minimum hazardous chemicals. Green chemistry principles started to be required for pharmaceutical quality control on a daily basis [14,15,16].
Regarding this investigation, a recently studied veterinary benzenedisulfonamide derivative with strong fasciolicidal activity is called Clorsulon (CLR). Through specific antagonistic interactions with fluke phosphoglycerate mutase and kinase, it disrupts the energy metabolism. Cattle internal and external parasites are commonly treated and controlled with the use of CLR injectable medication as a parasiticide [17]. Fig. S1 supplementary data shows CLR, molecular weight 380.66 g mol−1, chemical structure, and formula [C8H8Cl3N3O4S2]. By fermenting Streptomyces cyano-griseus, Moxidectin (MOX), a semisynthetic derivative of nemadectin, is produced. MOX’s molecular weight is 639.819 g mol−1 and its formula is [C37H53NO8], revealed in Fig. S1 supplementary data. MOX is an anthelmintic medication used to prevent and treat intestinal and heartworms. It works by killing parasitic worms, or helminths [18]. It is present in medications administered for sheep, cattle, horses, dogs, and cats. Depending on the treatment, there are several ways to apply moxidectin: oral, topical, and injectable preparations. Moxisulon Solution for Injection is the brand name used to distribute the CLR and MOX injectable solution. Fig. S2 supplementary data shows the UV spectrum for both CLR and MOX employing a spectrophotometric detector.
Recently, quite a few HPLC studies for the CLR assay in their dosage form [17, 19,20,21]. Two HPLC techniques, one for CLR assay results in milk [22], the second in the kidney of cattle [23], and the UPLC method for CLR assay results in commercial bulk drug substance batches [24]. A UV-visible spectrophotometric method was used for the assay of CLR in its dosage form [25]. MOX was determined in human plasma using HPLC study [18], while in serum [26], in cattle tissues [27] using LC-MS, and in cattle hair by LC–MS/MS [28]. UV-spectrophotometric method used for the assay of MOX [29], spectrofluorimetric method [30], and matrix solid-phase dispersion technique for the determination of MOX in bovine tissues [31]. The LC-MS instrument’s high cost and high energy consumption prevented it from being used on a daily basis for CLR and MOX tests in pharmaceutical companies’ quality control [32]. Actually, as of the writing of this text, no stability-indicating HPLC-DAD procedures have been found that have been published for the analysis of CLR and MOX as a mixture simultaneously, either with or without chemical derivatization.
Because of its applicability and effectiveness, HPLC is the best practical method from chromatography in analyzing medications when contaminants and degradates coexist in dosage forms [33]. The main goal of this study is to use green chemistry procedures to deliver the first straightforward, environmentally friendly HPLC-DAD for CLR and MOX as a therapeutically effective mixture [34] together with their degradation byproducts. Furthermore, the objective was extended to suggest optimal storage settings for CLR and MOX mixtures in veterinary pharmacies by utilizing the stability-representative method’s results. Finally, the two automated and current greenness assessment tools, AGREE and GAPI, have been used to evaluate the method’s effectiveness in terms of environmentally friendly aspects.
Experimental
Reagents, parenteral formulation, and pure chemicals
The Drug Authority in Egypt (El-Maadi, Egypt) provided friendly supplies of Clorsulon (CLR), Moxidectin (MOX), a genuine, pristine impurity proportion less than 1.98%, and MOXISULON Solution for Injection (Merial S.A.S., France). A sterilized injectable solution that is prepared for usage containing CLR and MOX is called MOXISULON Solution for Injection. A milliliter (mL) of MOXISULON solution contains 400 mg of formal glycerol, 300 mg of propylene glycol, 100 mg of CLR, and 10 mg of MOX.
The chromatographic analysis employed gradient-grade acetonitrile and methanol content of at least 99.9% for each, which were purchased from the German company Sigma Aldrich Chemie GmbH in Steinheim. Acetic acid was purchased from El Nasr Pharm. (Abu-Zabal, Egypt). Internal preparation of deionized water was accomplished by using the (arium®) mini Ultrapure Water System (Geottingen, Germany). Prior to chromatographic analysis, all liquid solutions underwent filtering and degassing. In the filtration process, (Tisch®) nylon membrane filters with a pore size of 5.0 μm, 47.0 mm, were employed.
Instrumentation
1200 Series Agilent with G1316A TCC Thermostatted Column Compartment, G1367C HiP-ALS SL autosampler, online degasser (G1322A), quadruple gradient pump (G1312A), and temperature controller (G1316A) were used. For chemical detection and quantification, the UV detector (G1315B)’s chem station software was programmed. Additionally, Memmert Co.‘s (Schwabach, Germany) equipment was used for the sonication procedure.
Procedure and chromatographic conditions
The optimization of the HPLC-DAD method was attempted multiple times. One variable at a time optimization was used to change a variety of factors, including composition, speed of the mobile phase, detection wavelength as shown in Fig. 1 and S3. The ideal parameters for producing the best chromatograms with regard to ultimate peak area, lowest backdrop noise, and sharpness for the main peaks for CLR and MOX are shown by Table 1.
Approaches for preparing standard and pharmaceutical solutions
A. Preparation of standard solutions
The primary standard solution was prepared by adding 70 mL of liquid mobile solution to 100 mL glass flasks after 200 and 20 mg of pure powder CLR and MOX were carefully added, one at a time, respectively. Each glass flask was then sonicated for thirty minutes. The glass flasks were then filled with the liquid mobile solution until the final volume reached 100 mL, giving concentrations of 2000.00 and 200.00 µg mL−1 for CLR and MOX, respectively. The secondary standard solution was prepared by transferring 10 mL from the primary standard solution to a 50 mL volumetric flask and then diluted to volume using the mobile phase, giving concentrations of 400.00 and 40.00 µg mL−1 for CLR and MOX, respectively.
B. MOXISULON prepared liquid for injection
One milliliter of the injectable MOXISULON solution, which contained 100 mg of CLR and 10 mg of MOX, was carefully added to a 50 mL flask to prepare the primary standard solution, then 30 mL of liquid mobile solution was added and the glass flask was then sonicated for thirty minutes. The glass flasks were then filled with the liquid mobile solution until the final volume reached 50 mL giving concentration (2000.00, and 200.00 µg mL−1) for CLR and MOX, respectively. 10 mL from (the primary standard solution) was transferred to 50 mL volumetric flask, then diluted to volume using the mobile phase giving concentration (400.00, and 40.00 µg mL−1) for CLR and MOX, respectively.
Procedures for the analysis of the CLR and MOX stability study
In accordance with ICH guidelines, methodologies for HPLC degrading stabilization [35] were carried out to analyze CLR and MOX in the MOXISULON formula in various stressed settings. Table 2 provided complete descriptions of the various deterioration environments. Additionally, the blank solution, composed of all formulation components except the drugs under investigation, was examined for comparative purposes. It was set up by adding all of the previously listed inactive additives, expressed as a percentage, to the MOXISULON solution without CLR or MOX.
In terms of light treatment, from the previously prepared solutions in Sect. 2.4.A, concentrations of 800.00 and 80.00 µg mL−1 for CLR and MOX, respectively, were exposed to either UV radiation for 12 h or direct sunshine for two days before being filtered via a syringe filter. The studied drug liquids underwent an eight-hour thermally regulated water bath at 80 °C as part of the heat-based procedure. The studied drug solution was then quickly filtered through sterile syringe filters after cooling.
Using a 50 mL volumetric flask, 200 mg of CLR and 20 mg of MOX were mixed with 5 mL of 1 N HCl and NaOH, each one at a time for the acid and basic assessments. After that, the bottle was heated for 60 min in a thermally controlled water bath set at 80 °C, and 1 N HCl or NaOH was used to maintain the solution’s pH at 7.0 ± 0.1. The final size was then achieved using the mobile phase and sonicating it for thirty minutes, giving a primary solution with concentrations of 4000.00 and 400.00 µg mL−1 for CLR and MOX, respectively. 10 mL from the primary solution was transferred to a 50 mL volumetric flask, then diluted to volume using the mobile phase, giving a secondary solution with concentrations of 800.00 and 80.00 µg mL−1 for CLR and MOX, respectively. Additionally, in H2O2 treatment, a 50 mL volumetric bottle containing 200 mg of CLR, 20 mg of MOX, and 5 mL of H2O2 (0.50%) was placed in a thermally adjusted water bath set to 80 °C for 60 min. After that, the pH of the solution was adjusted to 7.0 ± 0.1 using 1 N NaOH, and 35 mL of the liquid system was added. The bottle was then shaken for thirty minutes using a sonicator. Finally, the liquid solution was diluted like in the previous manner to reach final concentrations (800.00 and 80.00 µg mL−1) for CLR and MOX, respectively.
Evaluation of the validity of the HPLC approach
The International Conference of Harmonization (ICH) guidelines for endorsement of verification of techniques are considered regarding many parameters, including efficient linear manner, precision, specificity, accuracy, robustness, ruggedness, limits of detection, and limit of quantitation [36].
Evaluation of the expected HPLC method’s environmental benefits using the AGREE and GAPI tools
The automated software AGREE [37] and GAPI [38] were mostly utilized for method greenness assessments in order to verify the risks to the environment and analysts. We looked at the created pictograms for evaluation and methodologies, whereas subdivisions in green stand for totally safe analytical processes. Numerous publications have proven the effectiveness and dependability of the aforementioned tools [39].
Results and discussions
Even though there aren’t many published LC-MS and LC-MS/MS approaches for the CLR and MOX tests in pharmacokinetics investigations, high-performance liquid chromatography is the recommended apparatus for analyzing medicines in authentic shape, together with medicinal products, as the introduction describes due to concerns about energy and cost consumption [32]. Furthermore, the HPLC method’s speed and environmental friendliness significantly increase its dependability and suitability for everyday drug assays [40]. The daily quality control in pharmaceutical firms greatly depends on the improvement of a straightforward and eco-friendly stability-proving technique for a combination of CLR and MOX.
Outcomes of accelerated degradation studies
Figure 2 showed the collective chromatograms for the CLR and MOX mixture in various intensive degradation conditions, illustrating how well the approach separates the two compounds from potential degradates. Besides, the chromatograms for degradation studies were also displayed in their original state in supplementary file; whereas figure S3 illustrates the chromatogram of pure standard of the studied drugs, figure S4 illustrates a blank sample, figure S5 illustrates the studied drugs under photo degradation, figure S6 illustrates the studied drugs under thermal degradation, figure S7 illustrates the studied drugs under acidic degradation, figure S8 illustrates the studied drugs under basic degradation, and figure S9 illustrates the studied drugs under oxidative degradation. In conclusion, there was a clear separation between the peaks of CLR and MOX from those of other degradates. Table 1 lists the minor degradation products for the different demanding conditions. The minor peaks had retention durations of 3.10 and 3.50 min, but the CLR and MOX peaks had retention durations of 1.43 and 6.05 min. Table 2 displays the ratios of degradations. For CLR, the oxidative environment showed the highest degradation [about 27.56%], while the light degradation setting showed the lowest degradation [about 0.91%]. Likewise, for MOX, the acid environment showed the highest degradation [about 36.65%], while the light degradation setting showed the lowest degradation [about 1.10%]. One little peak emerged at 3.50 min as a result of the acidic degradations of the materials. The acidic degradations caused one small peak to appear at 3.50 min. The degradates in the basic and oxidative degradations were identical, giving only one peak that appeared at 3.10 min, and. There were no peaks for either photodegradation or heat-induced deterioration. The maximum observed retention times of 1.43 and 6.05 were measured over the whole 9-minute run to ensure that no additional small peaks were eluted from the column. The chromatographic conditions stated above, as reported in Table 1, enable rapid and efficient separation. The innovative study offers recommendations for CLR and MOX storage, highlighting how crucial it is to shield it from light and heat and the significance of using other drugs with caution, especially those that are oxidative, basic, or acidic.
HPLC chromatograms of the stability results of CLR (800 µg mL−1) and MOX (80 µg mL−1); the samples were subjected to (a) photodegradation, (b) thermal degradation, (c) acidic degradation, (d) basic degradation, and (e) oxidative degradation; the stationary phase was C-18, mobile phase of acetonitrile: methanol: water: acetic acid (56: 36: 7.5: 0.50 by volume), and flow rate 2 mL min−1 at 254 nm
Examining optimization for the HPLC technique
The C8 and cyanogen columns were eliminated from the selection process for the stationary phase. Conversely, as shown in Fig. 1, Supelcosil C18 (4.6 mm x 15 cm, 5 µ) produced acceptable resolution. Since CLR and MOX are basic in nature (pKa = 9.61 and 12.80 at 25 °C, respectively), the cyano column was disregarded. Furthermore, the stability and chromatographic behaviour of CLR and MOX may be impacted by the abundance of polar functional groups in MOX, such as secondary and tertiary amine groups, and the abundance of sulfoxide and hydroxyl groups in CLR. These functional groups may interact or establish hydrogen bonds on a cyano column, which could affect retention and separation. When the C8 column was compared to the C18 column, a wider peak was produced. As a result, C18 was chosen based on its peak form and was anticipated to have good degradation product resolution.
Our first preference for the mobile phase was acetonitrile, methanol, and water from an environmentally friendly standpoint. For the watery portion, a tiny amount of acetic acid was added. Because CLR and MOX are basic, adding acetic acid (pH = 4 ± 0.05) to the mobile system was also necessary to prevent peak tailing. The best mobile system components were also tested with several buffer types, and acetic acid alone was suggested as the pH adjuster. Furthermore, in order to improve peak uniformity, and resolution parameters an aqueous-free solvents, such as methanol or acetonitrile, were required especially for MOX which has high molecular weight and low solubility in water (0.50 mg/L) make its separation a challenging point [18]. Many non-aqueous solvents were tested, with ethanol being the most environmentally friendly option compared to acetonitrile and methanol. The chromatogram that was created for the ethanol case was insufficient. Furthermore, the researchers did their best to increase the ratio of water to organic solvents. However, inconvenient results were achieved in terms of peak symmetry and resolution. Using acetonitrile, methanol, water, and acetic acid (56: 36: 7.5: 0.50 by volume) as a mobile system, the best chromatogram was finally obtained. The wavelength of 254 nm was chosen for the DAD detector setup because it increases the sensitivity of the approach by coincidentally matching the λmax for both MOX and CLR.
To produce the best chromatogram in the shortest amount of time, numerous attempts were undertaken to optimize the flow rate. The retention periods for CLR and MOX were comparatively lengthy when operating at 0.5 mL min−1, free from degradation product interference. However, when operating at 1.0 mL min−1, the resolution was insufficient. The best results were obtained when the flow rate was set at 2.0 mL min−1. There were no other adjustments made to the ambient temperature.
Examining HPLC-DAD validation implications
To guarantee that the novel HPLC has a suitable linear scope and quantitation limit in addition to being robust, accurate, precise, specific, rugged, and reliable, the International Conference on Harmonization (ICH) practices were implemented [36]. Table 3 compiles all of the validation factor data, which were all satisfactory and in line with ICH principles. The supplemental PDF file S1 contains demonstrations of the comprehensive chromatograms for validation items. Figure S3 represented the HPLC chromatogram of pure standards for CLR and MOX using the aforementioned liquid system at 254 nm, while Figure S4 referred to the HPLC chromatogram of the blank sample. Therefore, the method specificity was concluded.
As shown in Table 4, five concentration values of each of the pure standards for CLR and MOX were chosen in order to examine linearity and create the regression equation. For CLR, the estimated linearity varied from 400.00 to 1200.00 µg mL−1, while for MOX, it was between 40.00 and 120.00 µg mL−1. Concerning the linear equation adaptations that were produced for each of CLR and MOX, the correlation coefficient value was close to unity. Though the CLR’s calibrating equation was [Y = 1.3047 X + 2.0172], r = 0.99988, it was [Y = 1.5447 X + 0.6618] for MOX and r = 0.99984. A complete positive relation between the variables CLR and MOX concentrations and their observed peak areas is indicated by correlation coefficient r values, which are close to unity. Additionally, as previously mentioned, 100 mg of CLR and 10 mg of MOX are included in each mL of the MOXISULON veterinary solution. With the use of the liquid mobile phase and successive dilutions, these can be measured effectively.
In order to test the accuracy of the procedure, Standard CLR and MOX are added to a blank sample that is made up of every component of the formulation, at predetermined amounts to spike the samples. The accuracy was assessed using a total of three tests at an identical quantity, and the concentrations’ recovery percentages were computed. The assessment is done at 50%, 100%, and 150% of the middle concentration. Different concentrations are measured: for CLR, the middle concentration is 800.00 µg mL−1, it is comparable to (400.00, 800.00, and 1200.00 µg mL−1), correspondingly. Similarly, for MOX, the middle concentration is 80.00 µg mL−1, it is comparable to (40.00, 80.00, and 120.00 µg mL−1), correspondingly. Predicted on the estimated recovery rates for specified concentration ± SD, the accuracy was evaluated; 100% ± 2 should be the accepted threshold. The correctness of the expected HPLC-DAD method was demonstrated by the recovery percentages (99.99%, 99.32%, and 100.12%) for CLR and (98.93%, 98.64%, and 101.16%) for MOX that were shown in Table 5. The inclusion of the three various concentrations stated above demonstrates that the procedure is inclusive of accuracy at each locus in the range of linearity.
To evaluate technique precision, six measurements were performed again for the same concentrations of CLR (800.00 µg mL−1) and MOX (80.00 µg mL−1) on the same day. The statistical test of relative standard deviation was used to evaluate the precision. The repeatability of the HPLC-DAD method was determined by RSD using Microsoft Excel computations, and the results were 0.15% and 0.37% for CLR and MOX, respectively. These results are shown in Table 6. A smaller RSD generally denotes a more precise approach.
By contrasting the retention durations and peak areas for the CLR and MOX standards, injection solutions, and placebo, the expected HPLC method’s specificity was considered. The predicted HPLC-DAD’s specificity is confirmed by Table 7’s comparable retention periods and peak area values. Supplemental PDF file S1 contained illustrations of the HPLC chromatograms for the injectable fluids, placebo, and CLR and MOX standards. Additionally, as can be shown in Fig. 2, HPLC chromatograms for CLR and MOX in various stressful settings demonstrated an excellent resolution for CLR and MOX peaks without interfering with the peaks of degradates. This validated the new RP-HPLC approach’s efficacy and selectivity.
Three replicates with different days and analysts were used to test the HPLC-DAD’s ruggedness, utilizing the same concentration (800.00 µg mL−1) for CLR and (80.00 µg mL−1) for MOX. The ruggedness of the expected HPLC-DAD approach was evaluated using the pooled relative standard deviation (SD) values (1.38 and 0.26%) for CLR and (1.64 and 0.66%) for MOX, which are listed in Tables 8 and 9. Pooled RSD is a statistical method that calculates the combined variability from different data sets. A lower pooled RSD value indicates better precision and ruggedness of the novel stability-illustrating method.
In order to verify the robustness of the approach, triple replicates at the same concentration (800.00 µg mL−1 for CLR and 80.00 µg mL−1 for MOX) were evaluated in two mobile systems with marginally different water/methyl alcohol ratios. Based on the pooled relative SD values (1.68%) for CLR and (1.31%) for MOX shown in Table 10, it was determined that the predicted HPLC-DAD method was robust.
Table 3 displays the values and formulae used in the computation of the limits of detection and quantitation. For CLR, the limits of detection (LOD) and limits of quantitation (LOQ) were 18.67 µg mL−1 and 56.58 µg mL−1, while for MOX, they were 2.17 µg mL−1 and 6.58 µg mL−1, respectively. The formulas [3.3 x σ / S] and [10 x σ / S], respectively [36], were used to calculate the values for LOD and LOQ. S denotes the calibration curve’s slope, while σ denotes the standard deviation of the responses of analytes.
To ascertain whether or not the innovative LC approach was appropriate to the analysis purpose, its suitability was further evaluated, achieving satisfactory results [41], shown in Table 11.
Review for the sustainability of the novel LC analytical approach
The reliability of the HPLC approach was assessed using 2 computational green meter methodologies: AGREE [37] and GAPI [38]. In the GAPI pictogram (Fig. 3), there were only three red subsets identified: stats 1, 7, and 15 show that offline testing techniques were used, acetonitrile was used as one component in the mobile phase, and that there was no waste management for the reagents used. The remaining thirteen parts, which were categorized as either green or yellow, accepted the established HPLC-DAD stability method. Moreover, the proposed stability method’s eco-friendly appraisal is approved by the AGREE pictogram (Fig. 4). Due to the position of the offline analytical device, only red subsection 3 was observed. Subdivision 7 exhibited an orange color due to a mobile liquid waste volume of approximately 18 mL per run, which is considered a moderate amount. Besides, subdivision 11 was also orange in color because of acetonitrile toxicity (56% of the mobile phase composition). With a combined score of 0.56, the other subdivisions, which varied from dark green to yellow due to a high percentage of acetonitrile in the mobile liquid phase, also demonstrated the relative greenness merit.
The amount of waste generated can be computed by multiplying the flow rate by 9 mL of solvent (a combination of acetonitrile, methanol, water, and acetic acid) for the entire run period. Section 7 of the AGREE tool classifies this volume as moderate. As a result, a light orange hue is generated, signifying a moderate amount of waste. For instance, the greenest solvents are methanol and water. We made every effort to stay away from acetonitrile. However, employing acetonitrile without it produced inaccurate chromatograms. Methanol mixed with acetonitrile is a compostable and (non-perfluoroalkyl, non-polyfluoroalkyl) fluid. It is crucial to remember that methanol and acetonitrile are dangerous substances even though they are not PFAS chemicals.
Furthermore, carbon footprint (kg CO2) is a crucial environmental and sustainable indicator. It might be determined via using the HEXAGON method and expressed as kg CO2 equivalent, as per the 2019 study by Ballester-Caudet and colleagues [42]. Supplementary data (page 3, assessment of carbon footprint environmental impact) contains an equation for the calculation and findings. The total amount of carbon dioxide equivalent left behind was 0.00741 kg. Using the HEXAGON method, the ultimate score is 0 out of 5 since the total calculated carbon footprint is less than 0.10. An eco-friendly method is one with a reduced carbon footprint score [15].
Without a doubt, the new HPLC-DAD stability method and its application to the MOXISULON injectable solution, when compared to the previously reported LC-MS bioanalytical methods, have the advantages of being greener, more economical, consuming less energy, and being simpler to implement, as shown in Table 12.
The innovative HPLC Approach’s benefits for the Pharmaceutical Sector
Considering the apparatus for HPLC is more affordable than that for LC-MS, the novel HPLC approach frequently serves as the initial cost-saving technique for the measurement of MOX and CLR. Furthermore, when the compounds are isolated from the related degradation products, the HPLC method—which is easier to use than the LC-MS method—is recommended for verifying efficacy and medication safety in frequent quality assurance analyses of CLR and MOX. Chromatography-grade liquids are generally easier to get, less expensive, and more widely available on the market than LC-MS-grade ones. Additional purification procedures are performed on LC-MS-grade solvents to reduce contaminants that may interfere with mass spectrometry analysis. In addition, the HPLC equipment uses about ten times less energy than the LC-MS one.
Suggestions regarding CLR and MOX Storage
Degradation ratio results shown in Table 2 indicate that strong basic or acidic medications and oxidants shouldn’t be combined with CLR and MOX in a co-formulated medication. It is best to keep CLR and MOX out of direct sunlight and heat.
Applicability and functionality of the method using the Blue Applicability Grade Index (BAGI) method
Manousi and her team [43] published the BAGI tool lately in 2023. In order to provide a pictogram and a score that demonstrate the applicability and efficiency of an analytical approach, BAGI takes 10 elements into account. To make the analytical process considered practical, it is recommended that the final score be higher than 60. The innovative HPLC method’s functioning and utility are indicated by its final score of 80.0 in Fig. 5.
Restrictions and potential futures
Since our main objective was to assess both CLR and MOX in sterile pharmaceutical solutions utilising an accessible and easy-to-use chromatographic apparatus, animal tissues and fluids were not used in the testing of the HPLC-DAD approach. Furthermore, because the LC-MS approach is able to identify at the nanogram scale, it may be the best method for evaluating CLR and MOX in animal tissues and fluids. While doing this research, our laboratory did not have access to LC-MS or GC-MS, which are suggested methods for determining the molecular weight of each degradate. Furthermore, more investigation is needed to clarify the mechanism of CLR and MOX degradation. In addition, other detectors like the refractive index, corona spray, and evaporative light scattering detector (ELSD) could be explored in comparative stability experiments to obtain a comprehensive view of their degradation profiles.
Conclusions
For the first time, the straightforward, reliable, sustainable, and consistency-suggestive LC approach was improved to evaluate Clorsulon and Moxidectin in pure and injectable solutions. Furthermore, a ratio of 0.91% in the light stimulus to 27.56% in the oxidative environment for CLR and 1.10% in the enticement of light and 36.65% within the acidic conditions for MOX was observed for the degradation of Clorsulon and Moxidectin in all severe conditions. In less than nine minutes, the peaks for Clorsulon and Moxidectin were clearly separated from the peaks for all other degradation products. Refrain from exposing Clorsulon and Moxidectin to heat or light, and use caution when co-formulated with basic, acidic, or oxidative medications. Finally, the new approach worked well for repeatable tests for Moxidectin and Clorsulon in pharmaceutical companies’ quality control laboratories.
Data availability
The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.
References
Sehrawat R, Maithani M, Singh R. Regulatory aspects in development of stability-indicating methods: a review. Chromatographia. 2010;72:1–6. https://doiorg.publicaciones.saludcastillayleon.es/10.1365/s10337-010-1612-z.
Chew YL, Khor MA, Lim YY. Choices of chromatographic methods as stability indicating assays for pharmaceutical products: a review. Heliyon. 2021;7:e06553. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.heliyon.2021.e06553.
Hegazy AM, Batubara AS, Abdelgawad MA, El-Sherbiny M, Ghoneim MM, Ahmed AM, Gamal M. Recommended and verified stability indicating GC–MS procedures for green separation of quaternary mixture of naphazoline, ephedrine, methylparaben, and naphazoline impurity. Microchem J. 2022;183:108058. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.microc.2022.108058.
Marzouk HM, Rezk MR, Gouda AS, Abdel-Megied AM. A novel stability-indicating HPLC-DAD method for determination of favipiravir, a potential antiviral drug for COVID-19 treatment; application to degradation kinetic studies and in-vitro dissolution profiling. Microchem J. 2022;172:106917. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.microc.2021.106917.
de la Guardia M, green analytical Chemistry SG. U. The concept of green analytical chemistry, Books. Google.Com. (2012). https://books.google.com/books?hl=en&lr=&id=a2cc97Y3SO8C&oi=fnd&pg=PA3&dq=Green+analytical+chemistry++M+de+la+Guardia+&ots=EwMhpGg3Sd&sig=S5Zv-1R82dErW9fCVouN1o1FYoI . Accessed 22 Jan 2023.
De La Guardia M, Garrigues S. Chap. 1: past, present and future of green analytical chemistry. RSC Green Chem Royal Soc Chem. 2020;1–18. https://doiorg.publicaciones.saludcastillayleon.es/10.1039/9781788016148-00001.
Płotka-Wasylka J, Fabjanowicz M, Kalinowska K, Namieśnik J. History and Milestones of Green Analytical Chemistry, in: 2019: pp. 1–17. https://doiorg.publicaciones.saludcastillayleon.es/10.1007/978-981-13-9105-7_1
Sajid M, Płotka-Wasylka J. Green analytical chemistry metrics: a review. Talanta. 2022;238:123046. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.talanta.2021.123046.
Arabi M, Ostovan A, Li J, Wang X, Zhang Z, Choo J, Chen L. Molecular imprinting: green perspectives and strategies. Adv Mater. 2021;33:2100543. https://doiorg.publicaciones.saludcastillayleon.es/10.1002/adma.202100543.
Ostovan A, Arabi M, Wang Y, Li J, Li B, Wang X, Chen L. Greenificated molecularly imprinted materials for Advanced Applications. Adv Mater. 2022;34:2203154. https://doiorg.publicaciones.saludcastillayleon.es/10.1002/adma.202203154.
Arabi M, Ostovan A, Bagheri AR, Guo X, Wang L, Li J, Wang X, Li B, Chen L. Strategies of molecular imprinting-based solid-phase extraction prior to chromatographic analysis. TrAC - Trends Anal Chem. 2020;128:115923. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.trac.2020.115923.
Arabi M, Ostovan A, Bagheri AR, Guo X, Li J, Ma J, Chen L. Hydrophilic molecularly imprinted nanospheres for the extraction of rhodamine B followed by HPLC analysis: a green approach and hazardous waste elimination. Talanta. 2020;215:120933. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.talanta.2020.120933.
Abdelgawad MA, Abdelaleem EA, Gamal M, Abourehab MAS, Abdelhamid NS. A new green approach for the reduction of consumed solvents and simultaneous quality control analysis of several pharmaceuticals using a fast and economic RP-HPLC method; a case study for a mixture of piracetam, ketoprofen and omeprazole drugs. RSC Adv. 2022;12:16301–9. https://doiorg.publicaciones.saludcastillayleon.es/10.1039/d2ra02395d.
Abdel-Moety EM, Rezk MR, Wadie M, Tantawy MA. A combined approach of green chemistry and quality-by-design for sustainable and robust analysis of two newly introduced pharmaceutical formulations treating benign prostate hyperplasia. Microchem J. 2021;160:105711. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.microc.2020.105711.
de Marco BA, Rechelo BS, Tótoli EG, Kogawa AC, Salgado HRN. Evolution of green chemistry and its multidimensional impacts: a review. Saudi Pharm J. 2019;27:1–8. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.jsps.2018.07.011.
Kannaiah KP, Sugumaran A, Chanduluru HK, Rathinam S. Environmental impact of greenness assessment tools in liquid chromatography – a review. Microchem J. 2021;170:106685. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.microc.2021.106685.
Nilusha Padivitage S, Adhikari AM, Rustum. Simultaneous determination of ivermectin, clorsulon and their related substances in an injectable finished product by a stability-indicating RP- HPLC method. J Pharm Biomed Anal. 2022;210:114580.
Alvinerie M, Sutra JF, Badri M, Galtier P. Determination of Moxidectin in plasma by high-performance liquid chromatography with automated solid-phase extraction and fluorescence detection. J Chromatogr B. 1995;674:119–24.
Vegad Kunjal L, Paranjape Dipty B, Shah Dhwani A, Patel Ekta D, Patel Yogesh K. Patel Kaushik R, development and validation of RP-HPLC method for simultaneous estimation of Ivermectin and Clorsulon in Ivercam injection. Indo Am J Pharm Res.2017:7(08).
Ali MM, Elbashir E, Abdalaziz MN. Development and Validation for HPLC Method of Assay of Lvermectin and Clorsulon in Combined Pharmaceutical Dosage Form. Int J Homeopathy Nat Med. 2017;3(6):45–55.
Ahmed S, Saad NS, Ismail M, Soliman HE, Zaazaa. Validated Stability-Indicating RP-HPLC Method for Simultaneous Determination of Clorsulon and Ivermectin Employing Plackett-Burman Experimental Design for Robustness Testing, Journal of AOAC INTERNATIONAL, Volume 99, Issue 2, 1 March 2016, Pages 571–578.
Frank J, Schenck R, Wagner W, Bargo. Determination of Clorsulon residues in milk using a solid-phase extraction cleanup and Liquid Chromatographic determination. J Liq Chromatogr, 16, 1993 - Issue 2.
Markus J, Sherma J, Method I. Liquid Chromatographic/Ultraviolet determination of Clorsulon residues in kidney tissues of cattle. J AOAC Int. September 1992;75(5):937-941.
Zhao D, Rustum AM. Determination of Clorsulon and its related substances in commercial bulk drug substance batches by a rapid ion-pair UPLC method. J Pharm Biomed Anal. 2024;239:115896.
Rajen K, Jignasa M. Simultaneous equation method for the estimation of Ivermectin and Clorsulon in their combined pharmaceutical dosage form by UV-visible spectrophotometry, Biology, 2014.
Rafaela C, Baptista MAM, Fernandes S, Gilaverte, Sonia CN, Queiroz MR, Assalin VL, Ferracini, Alda LG, Monteiro, Felix GR, Reyes. Determination of Moxidectin in serum by Liquid Chromatography-Tandem Mass Spectrometry and its application in pharmacokinetic study in lambs. J Braz Chem Soc. 2017;28(2):250–6.
Alisa Khunachak AR, Dacunha SJ, Stout. Liquid Chromatographic Determination of Moxidectin Residues in Cattle Tissues and Confirmation in Cattle Fat by Liquid Chromatography/Mass Spectrometry, Journal of AOAC INTERNATIONAL, Volume 76, Issue 6, 1 November 1993, Pages 1230–1235.
Subbarao N, Mupeksha M, Ashma S, Ashwini K, Batheja R, Sambasivarao P. Sensitive, Rapid Estimation of Moxidectin in cattle hair by LC–MS-MS, LCGC Supplements, 2016, 14, Issue 3, 27–31.
Palani Sathish Babu and K. Anand Babu and Deivasigamani Revathi and Krishnan Chitra, Method Development and Validation of Moxidectin in Synthetic Mixture Using UV-Spectrophotometry, Chemistry. 2015,102426265.
Patricia Esmeralda Vazquez–Quintal. Roger Iván Rodriguez–Vivas, David Munoz–Rodriguez, liquid–liquid extraction for the spectrofluorimetric determination of moxidectin or abamectin in bovine plasma. Chem Pap. 2022;76:7441–9.
Alvinerie M, Sutra JF, Capela D, Galtier P, Fernandez-Suarez A, Horne E. and M. O’Keeffe, Matrix Solid-phase Dispersion Technique for the Determination of Moxidectin in Bovine Tissues, Analyst, October 1996, Vol. 121 (1469–1472).
Al-Sanea MM, Gamal M. Critical analytical review: rare and recent applications of refractive index detector in HPLC chromatographic drug analysis. Microchem J. 2022;178:107339. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.microc.2022.107339.
Gamal M. Analytical review: analytical techniques for hyoscine N butyl bromide. Analyst. 2020;145:2025–37. https://doiorg.publicaciones.saludcastillayleon.es/10.1039/D0AN00076K.
Yazwinski TA, Tucker CA, Powell JG, Reynolds J, Johnson ZB, Moore JJ, Prewett C, Griffin D, Haller B. A field study comparing fecal egg count reduction, weight gain and product safety in stocker cattle treated with either moxidectin or ivermectin with clorsulon. The Bovine Practitioner; 2007. pp. 146–50.
Ali HM, Gamal M, Ghoneim MM, Mohammed Abd L, Elhalim. Quantitative Analysis of Abamectin, Albendazole, Levamisole HCl and closantel in Q-DRENCH oral suspension using a Stability-Indicating HPLC-DAD method. Molecules. 2022. https://doiorg.publicaciones.saludcastillayleon.es/10.3390/molecules27030764.
Swartz ME, Krull IS. Handbook of Analytical Validation, 2012. https://doiorg.publicaciones.saludcastillayleon.es/10.1201/b12039
Płotka-Wasylka J, Wojnowski W. Complementary green analytical procedure index (ComplexGAPI) and software. Green Chem. 2021;23:8657–65. https://doiorg.publicaciones.saludcastillayleon.es/10.1039/d1gc02318g.
Gamal M, Naguib IA, Panda DS, Abdallah FF. Comparative study of four greenness assessment tools for selection of greenest analytical method for assay of hyoscine: N -butyl bromide. Anal Methods. 2021. https://doiorg.publicaciones.saludcastillayleon.es/10.1039/d0ay02169e.
Abdelrahman MM. Green Analytical Chemistry Metrics and Life-Cycle Assessment Approach to Analytical Method Development. Green Chem. Anal. Sample Prep. Cham, Cham: Springer; 2022. pp. 29–99. https://doiorg.publicaciones.saludcastillayleon.es/10.1007/978-3-030-96534-1_2.
Kelani KM, Elzanfaly ES, Saad AS, Halim MK, El-Zeiny MB. Different greenness assessment perspectives for stability-indicating RP-HPLC method used for the assay of isoxsuprine hydrochloride and four nephrotoxic and hepatotoxic photothermal degradation products. Microchem J. 2021;171:106826. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/j.microc.2021.106826.
Wells M, Dantus M. Validation of chromatographic methods, in: Anal. Instrum. Handbook, Third Ed., 2004: pp. 1015–1033. https://doiorg.publicaciones.saludcastillayleon.es/10.1201/9781315118024-31
Ballester-Caudet A, Campíns-Falcó P, Pérez B, Sancho R, Lorente M, Sastre G, González C. A new tool for evaluating and/or selecting analytical methods: summarizing the information in a hexagon. TrAC Trends Anal Chem. 2019;118:538–47. https://doiorg.publicaciones.saludcastillayleon.es/10.1016/J.TRAC.2019.06.015.
Manousi N, Wojnowski W, Płotka-Wasylka J, Samanidou V. Blue applicability grade index (BAGI) and software: a new tool for the evaluation of method practicality. Green Chem. 2023;25:7598–604. https://doiorg.publicaciones.saludcastillayleon.es/10.1039/D3GC02347H.
Funding
This work was supported and funded by the Deanship of Scientific Research at Imam Mohammad Ibn Saud Islamic University (IMSIU) (grant number IMSIU-DDRSP2502).
Author information
Authors and Affiliations
Contributions
“M. A.:. Investigation, Data curation, resources, writing-original draft, writing-review, and editing. H.S. and F. K.: Methodology, funding acquisition, project administration, and editing. A. N.: Writing-review and editing. L. M.: Methodology, investigation, editing and reviewing. M. G.: Data curation, writing-review and editing, and project administration. All authors reviewed the manuscript.”
Corresponding author
Ethics declarations
Ethics approval and consent to participate
Not applicable.
Consent for publication
Not applicable.
Competing interests
The authors declare no competing interests.
Additional information
Publisher’s note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary Information
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License, which permits any non-commercial use, sharing, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if you modified the licensed material. You do not have permission under this licence to share adapted material derived from this article or parts of it. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by-nc-nd/4.0/.
About this article
Cite this article
Elsayed, M.A., Algethami, F.K., Alotaibi, A.N. et al. A stability-illustrating HPLC-DAD method for assessment of two veterinary anti-parasitic drugs: appraisal of the method’s greenness and blueness. BMC Chemistry 19, 35 (2025). https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s13065-025-01401-9
Received:
Accepted:
Published:
DOI: https://doiorg.publicaciones.saludcastillayleon.es/10.1186/s13065-025-01401-9