Skip to main content

Table 6 Using the RSD calculation, the precision of the estimated stability-indicating RP-HPLC method was assessed for six measurements at the same concentrations of CLR (800.00 µg mL−1) and MOX (80.00 µg mL−1)

From: A stability-illustrating HPLC-DAD method for assessment of two veterinary anti-parasitic drugs: appraisal of the method’s greenness and blueness

Number of measurements

Peak area

Recovery percent

Guidelines for acceptance [38]

CLR

MOX

CLR

MOX

Measurement no. 1

1049.399

123.912

100.35

100.81

 

Measurement no. 2

1047.867

123.050

100.20

100.11

Measurement no. 3

1048.650

123.153

100.28

100.19

Measurement no. 4

1050.972

123.941

100.50

100.83

Measurement no. 5

1050.216

124.058

100.43

100.93

Measurement no. 6

1052.002

124.042

100.60

100.91

Mean of peak area

1049.9

123.7

100.39

100.63

SD

1.5

0.5

0.15

0.37

RSD

0.15%

0.37%

0.15%

0.37%

RSD ≤ 2%