Table 4 Robustness study of the proposed method for the determination of 200 ng mL-1 VZN (n = 3)

From: Development and validation of a novel fluorometric approach utilizing flow injection analysis for the measurement of vilazodone: application to dosage form and spiked human plasma

Parameter

  

Recovery͙͙ (%)

Standard deviation

RSD (%)

Optimum condition

  

98.95

1.15

1.16

Flow rate

 

0.4

101.51

0.92

0.91

 

0.6

99.83

1.34

1.34

Percentage of acetonitrile (%)

 

58

98.72

1.44

1.45

 

62

100.59

0.74

0.74

pH

 

4.9

100.68

1.38

1.37

 

5.1

101.59

0.76

0.75

λ excitation (nm)

 

239

99.09

0.84

0.85

243

99.25

0.94

0.95

λ emission (nm)

 

484

99.55

1.07

1.08

488

98.48

1.52

1.55

  1. RSD = Relative standard deviation