Skip to main content

Table 3 Evaluation of the precision of the proposed flow injection method for the determination of VZN

From: Development and validation of a novel fluorometric approach utilizing flow injection analysis for the measurement of vilazodone: application to dosage form and spiked human plasma

Precision level

Added concentration (ng mL-1)

Measured concentration (ng mL-1)

Recovery*

(%)

Standard deviation

RSD (%)

Intra-day

(Day1, n = 3)

50

49.6

99.20

0.71

0.72

200

197.90

98.95

1.15

1.16

300

301.20

100.40

1.40

1.39

Inter-day

(Whole days, n = 9)

50

49.42

98.83

1.88

1.90

200

198.70

99.35

1.67

1.69

 

300

301.77

100.59

0.85

0.84

  1. * Number of determinations = 3
  2. RSD = Relative standard deviation