Skip to main content

Table 2 Cyclosporine analysis in real plasma samples by validated probe in optimal conditions (Condition: pH 9.0, ZIF-8 MOF concentration of 100 ΜL, and incubation time of 2-minutes)

From: Development of a metal-organic framework-based nanosensor for determination of cyclosporine in plasma samples

No.

Gender

Added (µg mL− 1)

Found (µg mL− 1)a

Recovery (%)b

1

Female

-

0.1

0.5

0.092 ± 0.008

0.194 ± 0.010

0.582 ± 0.009

-

102.1

97.9

2

Male

-

0.1

0.5

0.088 ± 0.005

0.188 ± 0.009

0.578 ± 0.009

-

100.8

98.1

3

Male

-

0.1

0.5

0.093 ± 0.004

0.195 ± 0.011

0.591 ± 0.012

-

101.7

99.4

4

Male

-

0.1

0.5

0.094 ± 0.007

0.197 ± 0.008

0.588 ± 0.007

-

102.7

98.6

5

Male

-

0.1

0.5

0.091 ± 0.004

0.192 ± 0.008

0.576 ± 0.010

-

101.7

97.1

  1. a average of three replication (n = 3) + standard deviation
  2. b Recovery (%) = [cyclosporine concentration in samples (after spiking– before spiking)/Added] × 100