Skip to main content

Table 2 Accuracy and Precision of the proposed method

From: A green and verified high-performance liquid chromatographic technique for the concurrent measurement of a few veterinary drug residues in milk

 

Conc. (µg/mL)

Intra-day (repeatability)

Inter-day (intermediate precision)

Found Conc. ± SD

Accuracya (% R)

Precisionb (% RSD)

Found Conc. ± SD

Accuracya (% R)

Precisionc (% RSD)

IMD

3

2.98 ± 0.003

99.48

0.112

2.98 ± 0.005

99.37

0.152

8

7.99 ± 0.032

99.94

0.398

7.98 ± 0.041

99.78

0.515

20

19.92 ± 0.018

99.62

0.092

19.96 ± 0.078

99.81

0.39

FNM

3

3.00 ± 0.028

100.31

0.949

3.01 ± 0.029

100.18

0.949

8

8.08 ± 0.009

100.35

0.107

8.07 ± 0.012

100.88

0.147

20

20.10 ± 0.117

100.49

0.582

19.95 ± 0.157

99.75

0.787

SDD

3

3.02 ± 0.05

100.81

0.178

3.02 ± 0.005

100.89

0.173

8

7.99 ± 0.016

99.82

0.195

8.00 ± 0.013

99.95

0.166

20

20.06 ± 0.028

100.32

0.14

20.22 ± 0.138

101.9

0.685

  1. aAverage of % R of three replicates determinations
  2. bAverage of % RSD of three replicates determinations in the same day
  3. cAverage of % RSD of three replicates determinations at three successive days