Skip to main content

Table 4 System suitability test results of the developed RP-HPLC method for determination of vit. C and vit. A simultaneously

From: Eco-friendly RP-HPLC method for simultaneous determination of water-soluble and fat-soluble vitamins in nano-formula and pharmaceutical dosage forms

Parameter

Vit. C

Vit. A

(RSD%) < 2

0.93

0.61

Resolution Rs > 2

-

8.69

Tailing factor T ≤ 2

1.30

1.14

Capacity factor K > 2

2.76

5.86

Plates Number N > 2000

4949

7730