Skip to main content

Table 1 Assay validation parameters and regression equations for determination of vit. C and vit. A by the proposed chromatographic method

From: Eco-friendly RP-HPLC method for simultaneous determination of water-soluble and fat-soluble vitamins in nano-formula and pharmaceutical dosage forms

Parameter

Vit. C

Vit. A

Linearity

1–60 µg.mL− 1

0.5–30 IU.mL− 1

Regression equation

y = 34486x– 26,475

y = 23988x + 2052.4

Correlation coefficient (r)

0.9997

0.9999

Accuracy (Mean ± RSD)

99.49% ± 1.58

100.26% ± 1.86

LOD

0.3 µg.mL− 1

0.15 IU.mL− 1

LOQ

1 µg.mL− 1

0.5 IU.mL− 1

Repeatability (1)

101.26% ± 0.91

98.27% ± 0.63

Reproducibility (2)

101.41% ± 0.69

98.66% ± 0.89

  1. (1) The intra-day (n = 18), Average of six different samples at 100% analyte concentrations repeated three times within day;
  2. (2) The inter-day (n = 18), Average of six different samples at 100% analyte concentrations in three successive days