Table 1 Regression parameters and validation sheet for determination of EDO standard solution by the proposed method

From: Functionalized novel carbon dots from bell pepper seeds for sustainable green Edoxaban quantification

Parameters

Microwave CDs

Plate CDs

Excitation wavelength

310 nm

310 nm

Emission Wavelength

409 nm

409 nm

Linearity range

0.80–20.00 µg/mL

0.80–20.00 µg/mL

Slope

14.41

22.31

Coefficient of determination (R2)

0.9996

0.9994

Standard deviation of the residuals (Sy/x)

2.30

3.84

Limit of detection (LOD) a

0.23 µg/mL

0.22 µg/mL

Limit of quantitation (LOQ) b

0.69 µg/mL

0.72 µg/mL

Mean %recovery c ± SD

100.11 ± 1.35

100.05 ± 1.61

Intra-day precision d

100.55 ± 0.86

98.55 ± 0.98

Inter-day precision e

101.31 ± 0.59

100.35 ± 1.15

  1. a LOD = 3.3σ/ S (σ is the standard deviation of the response and S the slope of the calibration curve)
  2. b LOQ = 10σ/ S (σ is the standard deviation of the response and S the slope of the calibration curve)
  3. c Average recoveries of six different concentrations
  4. d Average recoveries of three replicates for three different concentrations (0.80, 10.00, and 20.00 µg/mL), on the same day
  5. e Average recoveries of three replicates for three different concentrations (0.80, 10.00, and 20.00 µg/mL), on different days