Skip to main content

Table 7 Folic acid content under different conditions

From: Establishment and validation of HPLC methods for the determination of folic acid and parabens antimicrobial agents on folic acid oral solution

Conditions

Content (%)

Average (%)

RSD (%)

column temperatures

23℃

95.16

94.93

0.12

27℃

94.89

wavelength

278 nm

94.93

282 nm

94.98

flow rates

1.0 mL/min

94.95

1.4 mL/min

94.90

mobile phase ratio

0:100

94.92

3:97

94.68

pH

3.8

94.84

4.2

94.84

chromatographic columns

1

95.13

2

94.88

  1. RSD (%): Relative Standard Deviation = (SD/Mean) x 100