Skip to main content

Table 2 Validation data related to the suggested method for Tirofiban HCl quantification at MWCNT/CP Electrode

From: Electrochemical study of anti-platelets tirofiban HCl in dosage form using nanomaterials modified sensors: developed and green assessed by eco-scale and complex-GAPI approach

Parameters

(DPV)

Linear range (ng/mL)

27.00–745.00 (5.4 × 10–5 M-1.5 × 10−3 M)

Coefficient of determination (R2)

0.9995

Slope

0.0069

SD of slope

1.64

Intercept

0.5884

Accuracy (Mean% ± SD)

100.25 ± 1.64

Reproducibility (%RSD)

0.952

Repeatability (%RSD)

1.247

LOQ (ng/mL)

26.60 (5.4 × 10–5 M)

LOD (ng/mL)

15.50 (3.1 × 10–5 M)

S.E

0.41

%E

0.41

  1. Where SD is the standard deviation, RSD is the relative standard deviation, S.E is the standard error, %E is the percent error