Table 4 Comparison of the proposed methods with other techniques for analysis of Met and PS in urine

From: Extraction of methamphetamine and pseudoephedrine by modified graphene oxide solid phase extraction method coupled to HPLC-UV in urine sample

Method

Analyte

Concentration range (µg/L)

LOD/LLOQ

(µg/L)

EM-SPME3-GC-FID [45]

Methamphetamine

5-500

2.0

MIP-SPME4-CE [46]

Pseudoephedrine

5-500

1.1

DLLME-SFOP1-HPLC-UV [48]

Methamphetamine

10-3000

2

Head-space-SPME2-GC-MS [49]

Methamphetamine

0.5–1000

0.2

DSPE-HPLC-UV (this study)

Methamphetamine

250-12000

250

MSPE5-HPLC-UV (this study)

Methamphetamine and Pseudoephedrine

1000–15,000

1000

  1. 1Electromembrane solid-phase microextraction-GC-FID
  2. 2Molecularly imprinted polymer- solid-phase microextraction-CE
  3. 3Dispersive liquid–liquid microextraction - solidification of floating organic drop-HPLC-UV
  4. 4Headspace solid-phase microextraction- GC-MS
  5. 5Magnetic solid-phase extraction-HPLC-UV