Skip to main content

Table 7 Comparison of the present methods with reported LC-MS/MS method

From: Development and validation of HPLC-UV and LC-MS/MS methods for the quantitative determination of a novel aminothiazole in preclinical samples

Method

Linear range (ng/mL)

Accuracy (% recovery)

Precision (% CV)

Ref.

LC-MS/MS

50-1000

-

-

[32]

LC-MS/MS

1.25–1250

98.66-109.83

1.35–13.21

Present work

HPLC

2060–20,600

101.00-101.67

0.89–1.99

Present work