Skip to main content

Table 1 System suitability parameters of the proposed HPTLC-densitometric and HPLC–DAD methods

From: A novel stability-indicating chromatographic quantification of the antiparkinsonian drug safinamide in its pharmaceutical formulation employing HPTLC densitometry and ion-pair HPLC–DAD

Method

Parameter

Acid/Alk.Deg

Oxi. Deg

SAF

4-HBD

Reference value [41]

HPTLC-densitometry

Retardation factor (Rf) ± 0.02a

0.03

0.39

0.53

0.64

 

Capacity factor (k’)b

32.33

1.56

0.88

0.56

 

Selectivity (α)c

20.7

1.77

1.57

 > 1

Resolution (Rs)d

7.52

1.94

1.82

Rs > 1.5

Tailing factor (T)

1.00

1.50

1.00

1.00

T ≤ 2

HPLC–DAD

Parameter

4-HBD

Oxi Deg

Acid/Alk Deg

SAF

Reference value [42]

Retention time (min ± 0.2)

1.91

2.41

4.07

5.63

 

Selectivity (α)c

1.71

2.37

1.54

 > 1

Resolution (Rs)d

1.72

4.28

4.14

Rs > 1.5

Tailing factor (T)

1.00

1.33

1.00

1.00

T ≤ 2

Column efficiency (N)

5836

1486

2946

3167

N ˃ 2000

Height equivalent to theoretical plate HETP (cm/plate)

2.57 × 10–3

1.00 × 10–2

5.09 × 10–3

4.74 × 10–3

 
  1. aRetardation factor (Rf) = distance travelled by the analyte/distance travelled by the solvent front
  2. bCapacity factor (k’) = (1-Rf)/Rf
  3. cSelectivity (α) = k’2/k’1 calculated for each of two successive peaks
  4. dResolution (Rs) = Rf2Rf1/0.5 (w1 + w2), where Rf is the retardation factor, and w is the peak width calculated for each of two successive peaks for HPTLC and = 2(tRB–tRA)/(WB+WA) for HPLC