Skip to main content

Table 1 System suitability parameters of the proposed HPLC–DAD method

From: A sustainable HPLC method coupled with diode array detection for versatile quantification of telmisartan, chlorthalidone and amlodipine in a fixed-dose antihypertensive formulation and dissolution studies

Parameter

TEL

CHT

AML

Reference value [31]

Retention time (min ± 0.2)

4.3

5.3

8.5

 

Capacity Factor (k’)

1.05

1.52

3.05

 

Selectivity (α)*

1.45 2.01

 > 1

Resolution (Rs)*

3.46 4.92

Rs > 2

Tailing factor (T)

1.15

1.21

1.24

T ≤ 2

Column efficiency (N)

5088

6875

6281

N ˃ 2000

Height equivalent to theoretical plate HETP  (cm/plate)

0.0049

0.0036

0.0040

 
  1. *Selectivity and resolution factors are calculated between each two successive peaks