Table 1 The electrochemical validation parameters of the designed GCE sensors

From: Eco-conscious potentiometric sensing: a multiwalled carbon nanotube-based platform for tulathromycin monitoring in livestock products

Parameters

Sensor 1 (MWCNT-free ISM)

Sensor 2 (MWCNT/ISM)

Slope (mV/decade)a

19.98

20.22

Intercept (mV)

256.46

286.14

Correlation Coefficient (r)

1

1

Linearity Range (M)

1.0 × 10–6–1.0 × 10–3

1.0 × 10–7–1.0 × 10–3

Working pH Range

2–7

2–7

Response Time (Sec.)

18

8

Lifetime (Weeks)

3

7

Accuracy (Mean ± RSD%)b

99.84 ± 0.700

100.23 ± 0.576

Repeatabilityc (%)

1.037

1.011

Intermediate Precisionc (%)

1.619

1.486

Robustnessd (%)

1.264

0.986

LOQ (M)

1.0 × 10–6

1.0 × 10–7

LODe (M)

1.73 × 10–7

9.76 × 10–8

  1. aAverage of three measurements
  2. bMean ± RSD% of recoveries for five concentrations of TUL in triplicates
  3. c RSD% values for three concentration levels (1.0 × 10–4, 1.0 × 10–5, 1.0 × 10–6 M) of TUL, each repeated three times intraday for repeatability and interday for three consecutive days for intermediate precision
  4. dRSD% values for previous concentrations under minor pH variations
  5. eLOD (detection limit) was determined by intercepting the extrapolated lines of non-responsive and the Nernstian sections in the potential profile