S. No. | Chromatographic conditions | Observations | Reference figure |
---|---|---|---|
Different column | |||
1 | Column- Phenomenex Luna C 18 250 × 4.6 mm (10 µm) M.P: Acetonitrile and water (95:5) Temp: 25 °C Flow: 1 mL/min Inj. Volume: 3 µL Time: 10 min Diluents: Acetonitrile and water as 90:10 | Peak shape is adequate but a short Retention Time required | Figure 2 |
2 | Column: Phenomenex Luna C18 250 × 4.60 mm (5 µm) M.P: Acetonitrile and water (95:5) Temp: 25 °C Flow: 1 mL/min Inj. Volume: 10 µL Time: 10 min Diluents: Acetonitrile and water as 90:10 | Peak splitting (peak shape was not adequate) | |
3 | Column: Eclipse plus C18 100 × 4.60 mm (3.5 µm) M.P: Acetonitrile and water (95:5) Temp: 25 °C Flow: 0.5 mL/min Inj. Volume- 3 µL Time: 5 min Diluents: Acetonitrile and water as 90:10 | Peak shape was not adequate, tailing factor > 2 | |
4 | Column: ACE Excel C18 100 × 2.1 mm (2 µm) M.P: Acetonitrile and water (95:5) Temp: 25 °C Flow- 0.2 mL/min Inj. Volume: 3 µL Time: 5 min Diluents: Acetonitrile and water as 90:10 | Peak shape not adequate, theoretical plate < 2000 main peak starts with another peak, no clear separation | |
Different mobile phase with different flow rate | |||
5 | M.P: Acetonitrile and Methanol in 95:5 Column: Ace Excel Super C18 Temp: 25 °C Flow: 0.1 mL/min Inj. Volume: 3 µL Time: 5 min Diluents: Acetonitrile and water as 90:10 | Theoretical plates are too low (lot lesser than 2000) | Figure 3 |
6 | M.P –Acetonitrile and Methanol in 95:5 Column- Ace Excel Super C18 Temp- 25 °C Flow-0.3 mL/min Inj. Volume- 3 µL Time- 5 min Diluents- Acetonitrile and water as 90:10 | Retention Time is too high | |
7 | M.P: Acetonitrile and Methanol in 95:5 Column: Ace Excel Super C18 Temp: 25 °C Flow: 0.5 mL/min Inj. Volume: 3 µL Time: 5 min Diluents: Acetonitrile and water as 90:10 | Theoretical plates are too low (less than 2000) | |
8 | M.P: 0.1%TFA in Acetonitrile and methanol in ratio 95:5 Column: Ace Excel Super C18 Temp: 25 °C Flow: 0.1 mL/min Inj. Volume: 3 µL Time: 5 min Diluents: Acetonitrile and water as 90:10 | Peak not detected | Figure 4 |
9 | M.P: 0.1%TFA in Acetonitrile and methanol in ratio 95:5 Column: Ace Excel Super C18 Temp: 25 °C Flow: 0.3 mL/min Inj. Volume: 3 µL Time: 5 min Diluents: Acetonitrile and water as 90:10 | Theoretical plates are too low | |
Different temperature and flow rate | |||
10 | Column: ACE Excel Super C18 100 × 2.1 mm (2 µm) M.P: Acetonitrile and water (95:5) Temp: 45 °C Flow: 0.5 mL/min Inj. Volume- 3 µL Time: 5 min Diluents: Acetonitrile and water as 90:10 | Theoretical plates and resolution are too low | Figure 5 |
11 | Column: ACE Excel Super C18 100 × 2.1 mm (2µm) M.P: Acetonitrile and water ( 95:5) Temp: 45 °C Flow: 0.3 mL/min Inj. Volume: 3 µL Time- 5 min Diluents- Acetonitrile and water as 90:10 | Theoretical plate and resolution are too low | |
12 | Column- ACE Excel Super C18 100 × 2.1 mm (2µm) M.P –Acetonitrile and water ( 95:5) Temp: 45 °C Flow: 0.2 mL/min Inj. Volume: 3 µL Time: 5 min Diluents: Acetonitrile and water as 90:10 | Theoretical plates and resolution are good | |
13 | Column: ACE Excel Super C18 100 × 2.1 mm (2µm) M.P: Acetonitrile and water (95:5) Temp: 45 °C Flow: 0.1 mL/min Inj. Volume: 3 µL Time: 5 min Diluents: Acetonitrile and water as 90:10 | Theoretical plates and resolution are good but RT is not adequate | |
14 | Column: ACE Excel Super C18 100 × 2.1 mm (2µm) M.P: Acetonitrile and water (95:5) Temp: 25 °C Flow: 0.5 mL/min Inj. Volume: 3 µL Time: 5 min Diluents: Acetonitrile and water as 90:10 | Theoretical plates and resolution are too low | Figure 6 |
15 | Column: ACE Excel Super C18 100 × 2.1 mm (2 µm) M.P:Acetonitrile and water (95:5) Temp: 25 °C Flow: 0.3 mL/min Inj. Volume: 3 µL Time: 5 min Diluents: Acetonitrile and water as 90:10 | Theoretical plates are less than 2000 but resolution is good | |
16 | Column: ACE Excel Super C18 100 × 2.1 mm (2µm) M.P: Acetonitrile and water (95:5) Temp: 25 °C Flow: 0.2 mL/min Inj. Volume: 3 µL Time: 5 min Diluents: Acetonitrile and water as 90:10 | Theoretical plates and resolution are good and can be adopted in the present method | |
17 | Column: ACE Excel Super C18 100 × 2.1 mm (2µm) M.P: Acetonitrile and water (95:5) Temp: 25 °C Flow: 0.1 mL/min Inj. Volume: 3 µL Time: 5 min Diluents: Acetonitrile and water as 90:10 | Theoretical plates and resolution are good but RT is not adequate |